'''Total organic carbon''' ('''TOC''') is the amount of [[carbon]] bound in an [[organic compound]] and is often used as a non-specific indicator of [[water quality]] or cleanliness of pharmaceutical manufacturing equipment.
A typical analysis for TOC measures both the total carbon present as well as the [[Inorganic chemistry of carbon|inorganic carbon]] (IC). Subtracting the inorganic carbon from the total carbon yields TOC. Another common variant of TOC analysis involves removing the IC portion first and then measuring the leftover carbon. This method involves purging an acidified sample with carbon-free air or nitrogen prior to measurement, and so is more accurately called non-purgeable organic carbon (NPOC)<ref>Standard Methods for the Examination of Water and Wastewater, Method 5310A</ref>.
==TOC Analysis==
===Environmental===
Since the early 1970s, TOC has been recognized as an analytical technique to measure water quality during the drinking water purification process. TOC in source waters comes from decaying natural organic matter (NOM) and from synthetic sources. Humic acid, fuvic acid, amines, and urea are types of NOM. Detergents, pesticides, fertilizers, herbicides, industrial chemicals, and chlorinated organics are examples of synthetic sources<ref>Hendricks, David W., Water Treatment Unit Processes: Physical and Chemical, CRC Press, 2007, pp 44 - 62</ref>. Before source water is treated for disinfection, TOC provides an important role in quantifying the amount of NOM in the water source. In water treatment facilities, source water is subject to reaction with chloride containing disinfectants. When the raw water is chlorinated, active chlorine compounds (Cl<sub>2</sub>, HOCl, ClO<sup>-</sup>) react with NOM to produce chlorinated disinfection byproducts (DBPs). Many researchers have determined that higher levels of NOM in source water during the disinfection process will increase the amount of carcinogens in the processed drinking water. With passage of the [SDWA]http://www.epa.gov/ogwdw/sdwa/, TOC analysis emerged as a rapid and accurate alternative to the classical but lengthy biological oxygen demand (BOD) and chemical oxygen demand (COD) tests traditionally reserved for assessing the pollution potential of wastewaters. Today, Environmental Protection Agencies regulate the trace limits of DBPs in drinking water. Recent methods, such as USEPA method 415.3, D/DBP rule, regulate the amount of NOM to prevent the formation of DBPs in finished waters.
===Pharmaceutical===
Introduction of organic matter into water systems occurs not only from living organisms and from decaying matter in source water, but also from purification and distribution system materials. A relationship may exist between endotoxins, microbial growth, and the development of biofilms on pipeline walls and biofilm growth within pharmaceutical distribution systems. A correlation is believed to exist between TOC concentrations and the levels of endotoxins and microbes. Sustaining low TOC levels helps to control levels of endotoxins and microbes and thereby the development of biofilm growth. The United States Pharmacopoeia (USP), European Pharmacopoeia (EP) and Japanese Pharmacopoeia (JP) recognize TOC as a required test for purified water and water for injection (WFI). For this reason, TOC has found acceptance as a process control attribute in the biotechnology industry to monitor the performance of unit operations comprising purification and distribution systems. As many of these biotechnology operations include the preparation of medicines, the Food, Drug Administration (FDA) enacts numerous regulations to protect the health of the public and ensure the product quality is maintained. To make sure there is no cross contamination between product runs of different drugs various cleaning procedures are performed. TOC concentration levels are used to track the success of these cleaning validation procedures especially clean-in-place (CIP).
==Measurement==
To understand the analysis process better, some key basic terminologies should be understood and their relationships to one another (Figure 1).
'''Total Carbon (TC)''' – all the carbon in the sample, including both inorganic and organic carbon
'''Total Inorganic Carbon (TIC)''' – often referred to as inorganic carbon (IC), carbonate, bicarbonate, and dissolved carbon dioxide; a material derived from non-living sources.
'''Total Organic Carbon (TOC)''' – material derived from decaying vegetation, bacterial growth, and metabolic activities of living organisms or chemicals.
'''Non-Purgeable Organic Carbon (NPOC)''' – commonly referred to as TOC; organic carbon remaining in an acidified sample after purging the sample with gas.
'''Purgeable (volatile) Organic Carbon (POC)''' – organic carbon that has been removed from a neutral , or acidified sample by purging with an inert gas. These are the same compounds referred to as Volatile Organic Compounds (VOC) and usually determined by Purge and Trap Gas Chromatography.
'''Dissolved Organic Carbon (DOC)''' – organic carbon remaining in a sample after filtering the sample, typically using a 0.45 micrometer filter.
'''Suspended Organic Carbon''' – also called particulate organic carbon (PtOC); the carbon in particulate form that is too large to pass through a filter.
[[Image:TC Constituents.jpg|thumb|Figure 1: Flow Chart displaying the relationship of carbon components]]
Since all TOC analyzers only actually measure total carbon, TOC analysis always requires some accounting for the inorganic carbon that is always present. One analysis technique involves a two-stage process commonly referred to as TC-IC. It measures the amount of inorganic carbon (IC) evolved from an acidified aliquot of a sample and also the amount of total carbon (TC) present in the sample. TOC is calculated by subtraction of the IC value from the TC the sample. Another variant employs acidification of the sample to evolve carbon dioxide and measuring it as inorganic carbon (IC), then oxidizing and measuring the remaining non-purgeable organic carbon (NPOC). This is called TIC-NPOC analysis. A more common method directly measures TOC in the sample by again acidifying the sample it to a pH value of two or less to release the IC gas but in this case to the air not for measurement. The remaining non-purgeable CO<sub>2</sub> gas (NPOC)contained in the liquid aliquot is then oxidized releasing the gases. These gases are then sent to the detector for measurement.
Whether the analysis of TOC is by TC-IC or NPOC methods, it may be broken into three main stages:
#Acidification
#Oxidation
#Detection and Quantification
The first stage is acidification of the sample for the removal of the IC and POC gases. The release of these gases to the detector for measurement or to the air is dependent upon which type of analysis is of interest, the former for TC-IC and the latter for TOC (NPOC).
===Acidification===
The removal and venting of IC and POC gases from the liquid sample by acidification and sparging occurs in the following manner.
[[Image:Acidification and Purging Equation.jpg]]
===Oxidation===
The second stage is the oxidation of the carbon in the remaining sample in the form of carbon dioxide (CO<sub>2</sub>) and other gases. Modern TOC analyzers perform this oxidation step by one several processes:
# High Temperature Combustion
# High temperature catalytic (HTCO) oxidation
# Photo-oxidation alone
# Thermo-chemical oxidation
# Photo-chemical oxidation
# Electrolytic Oxidation
====High Temperature Combustion====
Prepared samples are combusted at 1,350<sup>o</sup> C in an oxygen rich atmosphere. All carbon present converts to carbon dioxide, flows through scrubber tubes to remove interferences such as chlorine gas, and water vapor, and the carbon dioxide is measured either by absorption into a strong base then weighed, or using an Infrared Detector<ref>http://www.tdi-bi.com/analytical_services/environmental/NOAA_methods/TOC.pdf</ref>. Most modern analyzers use non-dispersive infrared (NDIR) for detection of the carbon dioxide.
====High Temperature Catalytic Oxidation====
A manual or automated process injects the sample onto a platinum catalyst at 680<sup>o</sup> C in an oxygen rich atmosphere. The concentration of carbon dioxide generated is measured with a non-dispersive infrared (NDIR) detector.<ref>Sugimura and Suzuki, A High Temperature Catalytic Oxidation Method for the Determination of Non-Volatile Dissolved Organic carbon in Seawater by Direct Injection of a Liquid Sample, Marine Chemistry, 24 (1988) pp. 105 - 131</ref>
[[Image:Combustion Tube with Platinum Catalyst.jpg|thumb|Figure 2: A HTCO combustion tube packed with platinum catalyst]]
Oxidation of the sample is complete after injection into the furnace, turning oxidizable material in the sample into gaseous form. A carbon-free carrier gas transports the CO<sub>2</sub>, through a moisture trap and halide scrubbers to remove water vapor and halides from the gas stream before it reaches the detector. These substances can interfere with the detection of the CO<sub>2</sub> gas. The HTCO method may be useful in those applications where difficult to oxidize compounds, or high molecular weight organics, are present as it provides almost complete oxidation of organics including solids and particulates small enough to be injected into the furnace. The major drawback of HTCO analysis is its unstable baseline resulting from the gradual accumulation of nonvolatile residues within the combustion tube. These residues continuously change “TOC” background levels requiring continuous background correction. Because aqueous samples are injected directly into a very hot , usually quartz, furnace only small aliquots ( less than 2 milliliters and usually less than 400 micro-liters) of sample can be handled making the methods less sensitive than chemical oxidation methods capable of digesting as much as 10 times more sample. Also, the salt content of the samples do not combust, and so therefore, gradually build a residue inside the combustion tube eventually clogging the catalyst resulting in poor peak shapes, and degraded accuracy or precision.
====Photo-Oxidation (UV Light)====
In this oxidation scheme, ultra-violet light alone oxidizes the carbon within the sample to produce CO<sub>2</sub>. The UV oxidation method offers the most reliable, low maintenance method of analyzing TOC in ultra-pure waters.
====The UV/Chemical (Persulfate) Oxidation====
Like the photo-oxidation method, UV light is the oxidizer but the oxidation power of the reaction is magnified by the addition of a chemical oxidizer, which usually a persulfate compound. The mechanisms of the reactions are as follows:
[[Image:TOC Excitation and Oxidation Equations.jpg]]
The UV/chemical oxidation method offers a relatively low maintenance, high sensitivity method for a wide range of applications. However, there are oxidation limitations of this method. Limitations include the inaccuracies associated with the addition of any foreign substance into the analyte and samples with high amounts of particulates. By performing "System Blanks", which is to analyze then subtract the amount of carbon contributed by the chemical additive, helps lower inaccuracies but analyses in levels below 200ppb TOC are still difficult.
==== Thermo - Chemical (Persulfate) Oxidation ====
Also known as heated persulfate, the method utilizes the same free radical formation as UV persulfate oxidation except uses heat to magnify the oxidizing power of persulfate. Chemical oxidation of carbon with a strong oxidizer, such as persulfate, is highly efficient, and unlike UV, is not susceptible to lower recoveries caused by turbidity in samples. The analysis of system blanks, necessary in all chemical procedures, is especially necessary with heated persulfate TOC methods because the method is so sensitive that reagents cannot be prepared with carbon contents low enough to not be detected. Persulfate methods are used in the analysis of wastewater, drinking water, and pharmaceutical waters. When used in conjunction with sensitive NDIR detectors heated persulfate TOC instruments readily measure TOC at single digit parts per billion up to hundreds of parts per million depending on sample volumes.
===Detection and Quantification===
Accurate detection and quantification are the most vital components of the TOC analysis process. Conductivity and non-dispersive infrared (NDIR) are the two common detection methods used in modern TOC analyzers.
====Conductivity====
There are two types of conductivity detectors, direct and membrane. Direct conductivity provides an inexpensive and simple means of measuring CO<sub>2</sub>. This method has good oxidation of organics, uses no carrier gas, is good at the parts per billion (ppb) ranges, but has a very limited analytical range. Membrane conductivity relies upon the same technology as direct conductivity. Although it is more robust than direct conductivity, it suffers from slow analysis time. Both methods analyze sample conductivity before and after oxidization, attributing this differential measurement to the TOC of the sample. During the sample oxidization phase, CO<sub>2</sub> (directly related to the TOC in the sample) and other gases are formed. The dissolved CO<sub>2</sub> forms a weak acid, thereby changing the conductivity of the original sample proportionately to the TOC in the sample. Conductivity analyses assume that only CO2 is present within the solution.
As long as this holds true, then the TOC calculation by this differential measurement is valid. However, depending on the chemical species present in the sample and their individual products of oxidation, they may present either a positive or a negative interference to the actual TOC value, resulting in analytical error. Some of the interfering chemical species include Cl<sup>-</sup>, HCO<sub>3</sub><sup>-</sup>, SO<sub>3</sub><sup>2-</sup>, SO<sub>2</sub><sup>-</sup>, ClO<sub>2</sub><sup>-</sup>, and H<sup>+</sup>. Small changes in pH and temperature fluctuations also contribute to inaccuracy. Membrane conductivity analyzers have tried to improve upon the direct conductivity approach by incorporating the use of hydrophobic gas permeation membranes to allow a more “selective” passage of the dissolved CO<sub>2</sub> gas. While this has solved certain problems, membranes have their own particular limitations, such as with true selectivity, clogging and, more undetectably, they provide secondary sites for other chemical reactions, which are prone to display “false negatives,” a condition far more severe than “false positives” in critical applications. Micro leaks, flow problems, dead spots, microbial growth (blockage) are also potential problems. Most disconcerting is the inability of membrane methods to recover to operational performance after an overload or “spill” condition arises to over range the instrument, often taking hours before returning to reliable service and recalibration, just when accuracy of TOC analysis is most critical to operators for quality control.
====Non-dispersive infrared (NDIR)====
The non-dispersive infrared analysis (NDIR) method offers the only practical interference-free method for detecting CO<sub>2</sub> in TOC analysis. The principal advantage of using NDIR is that it directly and specifically measures the CO<sub>2</sub> generated by oxidation of the organic carbon in the oxidation reactor, rather than relying on a measurement of a secondary, corrected effect, such as used in conductivity measurements.
A traditional NDIR detector relies upon flow-through-cell technology, the oxidation product flows into and out of the detector continuously. A region of adsorption of infrared light specific to CO<sub>2</sub>, usually around 4.26 µm (2350 cm<sup>-1</sup>), is measured over time as the gas flows through the detector. The infrared adsorption spectra of CO<sub>2</sub> and other gases is shown in Figure 3. A second reference measurement that is non-specific to CO<sub>2</sub> is also taken and the differential result correlates to the CO<sub>2</sub> concentration in the detector at that moment. As the gas continues to flow into an out of the detector cell the sum of the measurements results in a peak that is integrated and correlated to the total CO<sub>2</sub> concentration in the sample aliquot.
[[Image:Atmosfaerisk spredning.gif|thumb|Figure 3: Plot of atmospheric transmittance in part of IR region showing CO<sub>2</sub> absorbing wavelengths]]
'''Recent Advances in NDIR Technology'''<br />
A new advance of NDIR technology is Static Pressurized Concentration (SPC).
The exit valve of the NDIR is closed to allow the detector to become pressurized. Once the gases in the detector have reached equilibrium, the concentration of the CO<sub>2</sub> is analyzed. This pressurization of the sample gas stream in the NDIR, a patent-pending technique, allows for increased sensitivity and precision by measuring the entirety of the oxidation products of the sample in one reading, compared to flow-through cell technology. The output signal is proportional to the concentration of CO<sub>2</sub> in the carrier gas, from the oxidation of the sample aliquot. UV/ Persulfate oxidation combined with NDIR detection provides good oxidation of organics, low instrument maintenance, good precision at ppb levels, relatively fast sample analysis time and easily accommodates multiple applications, including purified water (PW), water for injection (WFI), CIP, drinking water and ultra-pure water analyses.
[[Image:Fusion Total Organic Carbon Analyzer.jpg|thumb|Figure 4: A TOC Analyzer using SPC Technology]]
==Analyzers==
Virtually all TOC analyzers measure the [[CO2|CO<sub>2</sub>]] formed when organic carbon is oxidized and/or when inorganic carbon is acidified. [[Oxidation]] is performed either through [[Platinum_group#Applications|Pt-catalyzed]] combustion, by heated persulfate, or with a [[UV]]/[[Sodium persulfate|persulfate]] reactor. Once the CO<sub>2</sub> is formed, it is measured by a detector: either a conductivity cell (if the CO<sub>2</sub> is aqueous) or a non-dispersive infrared cell (after purging the aqueous CO<sub>2</sub> into the gaseous phase). Conductivity detection is only desirable in the lower TOC ranges in deionized waters, whereas NDIR detection excels in all TOC ranges. A variation described as [[Membrane Conductivity Detection]] can allow for measurement of TOC across a wide analytical range in both deionized and non-deionized water samples. Modern high-performance TOC instruments are capable of detecting carbon concentrations well below 1 µg/L (1 part per billion or ppb).
A total organic carbon analyzer determines the amount of carbon in a water sample. By acidifying the sample and flushing with nitrogen or helium the sample removes inorganic carbon, leaving only organic carbon sources for measurement. There are two types of analyzers. One uses combustion and the other chemical oxidation. This is used as a water purity test, as the presence of bacteria introduces organic carbon.
===Combustion===
In a combustion analyzer, half the sample is injected into a chamber where it is acidified, usually with phosphoric acid, to turn all of the inorganic carbon into carbon dioxide as per the following reaction:
CO<sub>2</sub> + H<sub>2</sub>O ↔ H<sub>2</sub>CO<sub>3</sub> ↔ H<sup>+</sup> + HCO<sub>3</sub><sup>-</sup> ↔ 2H<sup>+</sup> + CO<sub>3</sub><sup>2-</sup>
This is then sent to a detector for measurement. The other half of the sample is injected into a combustion chamber which is raised to between 600–700°C, some even up to 1200°C. Here, all the carbon reacts with oxygen, forming carbon dioxide. It's then flushed into a cooling chamber, and finally into the detector. Usually, the detector used is a non-dispersive infrared [[spectrophotometer]]. By finding the total inorganic carbon and subtracting it from the total carbon content, the amount of organic carbon is determined.
===Chemical Oxidation===
Chemical oxidation analyzers inject the sample into a chamber with phosphoric acid followed by persulfate. The analysis is separated into two steps. One removes inorganic carbon by acidification and purging. After removal of inorganic carbon persulfate is added and the sample is either heated or bombarded with UV light from a mercury vapor lamp. Free radicals form from the persulfate and react with any carbon available to form carbon dioxide. The carbon from both determination (steps) is either run through membranes which measure the [[Electrical conductivity|conductivity]] changes that result from the presence of varying amounts of carbon dioxide, or purged into and detected by a sensitive NDIR detector. Same as the combustion analyzer, the total carbon formed minus the inorganic carbon gives a good estimate of the total organic carbon in the sample.
This method is often used in online applications because of its low maintenance requirements.
For example the online Biotector which is the most modern application of this method.
==Applications==
TOC is the first chemical analysis to be carried out on potential petroleum source rock in [[oil exploration]]. It is very important in detecting contaminants in drinking water, cooling water, water used in semiconductor manufacturing, and water for pharmaceutical use. Analysis may be made either as an online continuous measurement or a lab-based measurement.
TOC detection is an important measurement because of the effects it may have on the environment, human health, and manufacturing processes. TOC is a highly sensitive, non-specific measurement of all organics present in a sample. It, therefore, can be used to regulate the organic chemical discharge to the environment in a manufacturing plant. In addition, low TOC can confirm the absence of potentially harmful organic chemicals in water used to manufacture pharmaceutical products. TOC is also of interest in the field of potable [[water purification#Tertiary treatment|water purification]] due to disinfection of byproducts. Inorganic carbon poses little to no threat.
==Walkley-Back modified method ==
{{Unreferenced-section|date=March 2007}}
===Calculation===
In the Total Organic Carbon determination to obtain the percent carbon content from soil, first we have to standardize the titrant solution (FeSO<sub>4</sub>•7H<sub>2</sub>O) before the sample analysis are made, as result we obtain some data which have to be reduced in order to obtain the results we need, and to do this, we use the next equations:
Titrant normality equation:
Eq.1<math>N_2=\frac{N_1V_1}{V_2}</math>
where:<br/>
N<sub>1</sub>: K<sub>2</sub>Cr<sub>2</sub>O<sub>7</sub> [[Molality#Normality|normality]]<br/>
V<sub>1</sub>: K<sub>2</sub>Cr<sub>2</sub>O<sub>7</sub> volume (mL)<br/>
V<sub>2</sub>: FeSO<sub>4</sub> volume (mL)<br/>
Organic carbon percentage:
Eq.2<math>%Carbon=\frac{(A-B)*0.3*1.33}{C}</math>
where:<br/>
A: [[meq]] K<sub>2</sub>Cr<sub>2</sub>O<sub>7</sub> = (mL K<sub>2</sub>Cr<sub>2</sub>O<sub>7</sub> x N K<sub>2</sub>Cr<sub>2</sub>O<sub>7</sub>)<br/>
B: meq FeSO<sub>4</sub>•7H<sub>2</sub>O = (mL FeSO<sub>4</sub>•7H<sub>2</sub>O x N FeSO<sub>4</sub>•7H<sub>2</sub>O)<br/>
C: grams of sample<br/>
0.3: Conversion factor to carbon weight, We have milliequivalents as result of the difference between A and B, and they need to be converted to carbon milliequivalents in order to get the units we need, for that it is necessary to do the next operation:
Eq.3<math>meq*(1eq/1000meq)*((1/4)molC/1eqC)*(12g/1molC)*100=0.3gC(%)</math>
The 0.3 conversion factor has units of carbon grams and involves the constant to convert a fraction to percent units; hence equation 2 does not have the factor 100.
Walkey-Black constant for sediments. 75% is the mean recuperation of carbon
in solids and sediments by using this method, that's why the final result has to be
multiplied by 1.33 in order to get the real value, this constant is not used when
determining carbon in KHP standard because almost all its carbon content is recovered.
==References==
<references/>
* Heggs, Eric T., Price, Edward K., Proffitt, Stephen R. “CO<sub>2</sub> Measurements For TOC Analysis Using Static Pressure Reading of an NDIR”. U.S. Patent Application
* “Visible/IR Lamps for NDIR Gas Sensors”. International Light Technologies
* “Theory and Operation of NDIR Detectors” Technical Note TN-169 rev 1 wh.02-02 RAE Systems Inc
* Wallace, B. “How to Gain Reliable, Fast, Precise and Accurate TOC Results for PW, WFI, UHPW or CIP Sample Analysis” Teledyne Tekmar.
* Infrared Spectra region plot, Wikipedia Encyclopedia, http://en.wikipedia.org/wiki/Infrared'''
* http://www.teledynetekmar.com
* Thermo Scientific, http://www.ankersmid.com
[[Category:Analytical chemistry]]
[[Category:Chemical oceanography]]
[[Category:Water quality indicators]]
[[cs:Celkový organický uhlík]]
[[de:Gesamter organischer Kohlenstoff]]
[[it:Total organic carbon]]
[[ja:全有機炭素]]
[[pl:Węgiel organiczny]]
[[sv:TOC]]
[[zh:总有机碳量]]